5025E Multifunction Calibrator 订购编号:
点击这里发送询价请详细填写一下内容,包括产品原产国、品牌完整名称以及您所能得到的所有信息,这将有助于我们准确迅速的报价。非常感谢!
Calibration at an Affordable Price
A precision multi-product calibrator designed for wide calibration workload coverage at an affordable cost. Accurately test and verify both traditional and new test equipment with a calibrator that meets both technical and budget requirements. The 5025E is ideal for companies requiring a calibrator that can be implemented quickly and easily, providing immediate return on investment. With increased capabilities it can be used as a multimeter calibrator and much more.
The 5025E is fitted with AC/DC voltage to 1050V, AC/DC current to 22A, digital frequency to 10MHz, decade resistance to 1GΩ, simulated variable resistance from 40Ω to 40kΩ, conductance, capacitance, thermocouple and PT100 simulation.
An optional enhanced performance pack increases the specification of certain functions for wider calibration coverage. It provides full range variable resistance to 120MΩ, extended temperature functions, extended capacitance, and increased ACV frequency bandwidth. Also available are oscilloscope and power calibration options.
In addition external adaptors are available for clamp meter calibration (up to 1100A) and optical tachometer calibration. Also available is a power amplifier which increases the 5025E's AC/DC voltage output drive capabilities to 100mA. This addition allows even the most demanding analogue instruments to be tested.
Multi Instrument and Device Calibration
The 5025E can calibrate various instruments and devices; bench and handheld multimeters, frequency meters, ohm meters, AC/DC millivoltmeters, thermocouple and PT100 meters, clamp meters, and many other measurement devices. By adding internal options the 5025E can calibrate oscilloscopes, timer/counters, power meters and more.
For more information see the 5025 Series Calibration Capabilities Chart showing the DUT coverage of the 5025E and 5025C calibrators.
Simple Operation
Functions and ranges are easily accessed from the front panel. Increase and decrease keys per digit, are used to quickly set the output value. Deviation control then enables the user to finely adjust the output value as a percentage (+/- 9.999%). All this information is shown on a clear, easy to read LED display.
Virtual Control Software
The 5025E is supplied as standard with Time Electronics' windows based Virtual Control software. It allows quick and direct input of any value required. The remote SCPI command is displayed to help users develop custom applications.
EasyCal Calibration Software
Connect the 5025E to a PC/Laptop (via RS-232 or USB) installed with Time Electronics' EasyCal software and automate the calibration process. This provides increased speed of calibration and consistency of results. Produce calibration certificates and reports to ISO 9001, ISO 17025, and other international quality standards.